| Technote # |
Title |
||||
| 41 | Getting Started Guide for STarStation | ||||
| Date Updated | Category | ||||
| 20-08-07 | System Operation | ||||
| Products | |||||
| [STarStation 2.0][STarStation 2.3] | |||||
|
For STarStation 2.3-link to PD 038-301 SS 2.3 operation guide- User Modules on Product page - Section Quick Guides For STarStation 2.0 Users please refer to the instructions below.
Note that IS version systems begin warming up immediately following power up whereas a warmup must be initiated by the user on 1.7 lx100 systems. This can be confirmed by the Laser Status Indicator on the instrument Status Bar. During warm-up the laser status displays the laser warming icon. 5. Select the Scripts Tab and select the default startup script. The startup script will warm the detectors, prime the system to remove air bubbles from the sheath buffer inlet tubing and perform an alcohol wash cycle to remove air bubbles from the cuvette.
-Eject the plate holder and place a ACS utility plate on the tray ensuring that the commands outlined in the script are satisfied R1 - Alcohol (require at least 3ml of 70% ethanol) Reservoir - Fill with Sheath Buffer. -Retract the tray. On IS version STarSystems the wash, soak, backflush, prime, drain and Fill and sanitize are able to proceed during laser warm-up, this is NOT possible on a 1.7 version STarSystems. NOTE: If the system is left idle for 4 hr, the lasers will automatically turn off. A 30 min warm-up period may again be required prior to reading an assay. 6. Once the Warm-up, Prime and Wash steps have completed verify that
the system pressure displayed on the instrument status bar reads between
6 - 9 psi. Note: the system sheath and air pressures will fall following
the completion of an instrument command when the compressor is disengaged.
Confirm that the instrument detectors are warm the Laser State icon should
display Probe Height Adjustment- See Technote #40 System Calibration It is recommended that the system be calibrated at least once a week and whenever the dCal Temperature displayed on the instrument status bar reads +/-3 0c. The STarSystem calibration status is displayed on the instrument status
bar
the individual state of the previous reporter and classifier calibrations and the date on which they were performed is also displayed. Note: the dCal temperature should only used as criteria for deciding whether system calibration is required after the machine has been warmed up and cleansed. Since the ambient temperature of the instrument changes following the system warm-up. 7. If you require a system calibration, ensure that the calibration bead bottles are vortexed prior to use. Important: Before vortexing, remove the calibration beads from 2-8°C storage and allow the beads to warm to room temperature. Stage2:Preparing the Assay Note: this step can be performed whilst the instrument is warming up. Before you can start collecting sample data you must enter the details for the assay you wish to perform into the STarStation Assay Manager. 1. Ensure you are logged into STarStation under your account name (Assay
information once added to the system is only available to the person that
entered it). Stage 3:Preparing the Plate/Worklist Note: this step can be performed whilst the instrument is warming up. 1. Ensure that you are logged-in to STarStation under your account. Stage4: Reading the Plate NOTE: Worklists must be created prior to reading a sample. The standard concentrations cannot be added / modified after the plate has been read. Therefore it is imperative that you ensure when creating and selecting Worklists that you assign time to check that the plate layout and assay described in the Worklist accurately reflect the physical setup of the microtitre plate containing your assay. 1. Check that the filter plate is flat. While pressing on one end of
the plate, observe the distance that the opposite end of the plate is
raised off a flat surface. If the distance is greater than 1 mm, transfer
all contents to a flat-bottom 96-well plate or another pre-wetted filter
plate and proceed with reading the assay (having ensured optimal probe
height). 7. After STarStation opens the template worklist, you should verify that
the correct worklist has been selected via inspecting that the number
of analytes the worklist has been setup to detect corresponds with either
the number of analyte tabs displayed in the data grid or the number of
bead regions displayed on the classifier plot. If there is a discrepancy
in the worklist setup, return to the worklist tab to review or edit the
worklist.
10. Press the Next Button and set the Platform Heater temperature if necessary.
13. Once data collection begins use the DD plot and Classifier plot to verify that adequate bead events are detected and that the distinct color coded beads are classified appropriately. When satisfied that sufficient beads are being counted we recommend leaving the system to collect data (modifications to gating or classification can be performed via the list mode data during analysis).
NOTE: It is possible to skip specific wells or analyze a well (sample) a second time using the pause acquisition function of the STarStation software. Note: If reading more than 1 plate, empty the waste and refill the sheath containers after each plate is run. We also recommend performing an alcohol flush followed by two washes with sheath fluid between each plate and warming the laser. Stage 5: Preparing STarSystem for Shutdown 1. When the assay reading is complete, select the Scripts tab from the STarStation instrument controls and select the default shutdown script.
2. The shutdown script will decontaminate the system and wash the salt-containing
sheath buffer from the sample needle and sample tubing reducing the potential
for salt precipitation. R2 - 3ml of 10% household bleach or hypochlorite solution. R3 - 2ml Distilled Water 4. Click the Start button to launch the script. Note: It is recommended that you wash with Distilled water (Soak) at least three times before shutting down the instruments. This step is crucial in preventing the precipitation of salts from the sheath buffer in both the sample tubing and sample probe.
|
|||||
| [ Close Window ] | |||||